EC Range |
0.00 to 29.99 vs./cm, 30.0 to 299.9 μS/cm, 300 to 2999 μS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm (absolute EC) |
EC Resolution |
0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.01 mS/cm, 0.1 mS/cm |
EC Accuracy |
±1% Of reading (±0.05 μS/cm or 1 digit, whichever is greater) |
EC Calibration |
1 Point offset calibration (0.00 μS/cm in air), 1 point slope calibration in EC standard 84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm |
TDS Range |
0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L (absolute TDS) |
TDS Resolution |
0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L |
TDS Accuracy |
±1% Of reading (±0.03 ppm or 1 digit, whichever is greater) |
TDS Calibration |
Through EC calibration |
EC/TDS Temperature Compensation |
Automatic 23 ºF to 212 ºF (-5 ºC to 100 ºC); NoTC – nonabsolute conductivity. |
Temperature Correction Coefficient |
0.00 to 6.00% / ºC (for EC and TDS only). Default value is 1.90% / ºC |
EC to TDS Conversion Factor |
0.40 to 0.80 |
Salinity (% NaCl) Range |
0.0 to 400.0% NaCl |
Salinity (% NaCl) Resolution |
0.1% NaCl |
Salinity (% NaCl) Accuracy |
±1% of reading |
Salinity (% NaCl) Calibration |
Single point with HI 7037L Standard |
Salinity (PSU) Range |
2.00 to 42.00 PSU |
Salinity (PSU) Resolution |
0.01 PSU |
Salinity (PSU) Accuracy |
±1% Of reading |
Salinity Range |
0.0 to 80.0 g/L |
Salinity Resolution (ppt) |
0.01 g/L |
Salinity Accuracy (ppt) |
±1% Of reading |
Temperature Range |
-4.0 to 248.0 °F (-20.0 to 120.0 ºC) |
Temperature Resolution |
0.18 ºF (0.1 ºC) |
Temperature Accuracy |
±0.9 ºF (±0.5 ºC) |
EC/TDS Probe |
HI763100 EC/Temperature probe |
Logging Type |
Manual Log on Demand (Max. 200 logs); Manual Log on Stability (Max. 200 logs); Interval Logging 100 lots, Max. 600 logs/lot |
Logging Memory |
Up to 1000 Records |
Connectivity |
1 Micro-USB port for charging and PC connectivity, 1 USB port for storage |
GLP |
yes |
Battery Type/Life |
Built-in rechargeable battery with up to 8 hours of continuous use |
Power Supply |
5 V DC Adapter (included) |
Environment |
(0 ºC to 50 °C) RH max 95% non-condensing |
Dimensions |
8 in x 5.5 in x 0.5 in (202 mm x 140 mm x 12.7 mm) |
Weight |
8.82 lb. (250 g) |