Scanning UV and Vis Spectrophotometer with Multiple Bandwidths

GT00WA00Z9

Description

Key Features

  • Large LCD screen with 65 thousand true color and support U disk data saving
  • Has an environmental Deuterium lamp, which prevents Ozone inhalation.
  • Data output: Data can be exported to a U disk (USB Output)
  • Print port: The Instrument can be connected to a PCL printer for printing A4 paper (USB Port)
  • Adopted high-class grating with a full light path design, to ensure the instrument has the super low stray light
  • With GLP self-check function, checks the wavelength accuracy and Photometry accuracy, can provide test report and power-off protection
  • Upgrade package can be downloaded from AOE web with the Serial Number, support U disk one-key upgrade

Technical Specifications

Optical System

 

Single Beam, Grating 1200 lines/mm
Wavelength Range 190-1100n
Spectral Bandwidth

 

0.5,1,2,4,5nm
Wavelength Accuracy ± 0.3nm
Wavelength Repeatability ≤ 0.2n
Photometric Accuracy ± 0.002A (0-0.5Abs), ± 0.004A (0.5-1.0Abs), ± 0.3% T (0-100% T)
Photometric Repeatability 0.001Abs (0-0.5Abs), 0.002Abs (0.5-1.0Abs), ≤ 0.1% T (0-100% T)
Stray Light ≤ 0.04% T @ 360nm; 220nm
Stability ± 0.001A / h @ 500nm
Baseline Flatness ± 0.0015A
Noise ± 0.0004Ab
Display 65 thousand true color 7 inch TFT LCD (480 *800
Photometric Mode

 

T, A, C, E
Photometric Range

 

0-200% T, -0.301-3.0A
Detector Silicon

 

Photodiode
Light Source

 

Deuterium Lamp, Tungsten Lamp
Input Membrane

 

Keypad
Output

 

USB2.0*2 Print and data output USB1.0 Connect PC

 

Compartment Optional 8 Auto Cell holder, Solid Sample Holder, Micro Cell Holder, 10-100mm (0.39- 3.9 in ) Cell Holder