Description
Features
- The high-frequency X-ray generator features high KV, high current, high frequency, and high voltage. Its core part is specifically designed to provide stable and reliable DC high voltage for the X-ray tube.
- Automatic exposure control of Ionization chamber (AEC)
- Internet real-time monitoring of the running status of equipment
- Remote intelligent upgrade and maintenance of Software
- High-precision time control of photography
- High-precision control of tube current Automatic fault protection and diagnosis
Technical Specifications
Max power output | 32 kW |
Nominal electric power (0.1s) | 32 kW |
X-ray tube voltage | Range: 40 kV to 125 kV
Pattern: Continuous adjustment Step length: 1 Deviation: ±10% |
X-ray tube current | Range: 32 mA to 400 mA
Pattern: Stepping adjustment Step length: According to R`10 numerical system Deviation: ±20% |
Load time | Range: 10 ms to 5000 ms
Pattern: Stepping adjustment Step length: According to R`10 numerical system Deviation: ± (10% + 1 ms) |
Current time product | Range: 0.32 mAs to 400 mAs
Pattern: Stepping adjustment Step length: According to R`10 numerical system Deviation: ± (10% + 0.2 mAs) |
Inverting frequency | ≥26 kHz |
Environmental Requirements
Temperature (working) | 14°F to 104°F (-10℃ to +40℃) | |
Temperature (transportation and storage) | -4°F to 131°F (-20°C to +55°C) | |
Humidity (working) | 30 % H.R. to75 % H.R. | Relative Humidity |
Humidity (transportation and storage) | 10 % H.R. to100 % H.R. | Relative Humidity |
Atmospheric pressure (working) | 70 kPa to 106 kPa | |
Atmospheric pressure (transportation and storage) | 50 kPa to 106 kPa |
Power Grid Requirements
Input power | 220 V to ±10% V | Sinusoidal alternating current |
Non sinusoidal distortion (V) | Less than 2% of its amplitude | |
Phase | Single phase, three-wire system | Independent safe grounding wire |
Mains frequency | 50 Hz or 60 Hz to ±1 Hz | |
Internal resistance of power source | <0.5 Ω | |
Power capacity | ≥15 kVA | |
Ground resistance | <4 Ω |