Description
Features
- Auto range function
- 32+ automatic measurements
- Advanced digital trigger system
- 14+ trigger function
- Serial triggering and decode
- Integrated USB Host/Device
- Convenient to communicate with PC
- Variety of SCPI
- RS232 port, LAN port. (optional)
- Stable and reliable quality
Technical Specifications
Bandwidth | 200 MHz |
Horizontal Sample Rate Range | 1 GS/s |
Horizontal Waveform Interpolation | (sin x)/x |
Horizontal Record Length | Maximum 64 K samples per single-channel |
Maximum 32 K samples per dual channel (4 K, 32 K optional) | |
Horizontal SEC or DIV Range | 2 ns/div to 100 s/div |
1, 2, 5 sequence | |
Sample Rate and Delay Time Accuracy | ±50 ppm |
Horizontal Delta Time Measurement Accuracy (Full Bandwidth) | Single-shot, Normal mode ± (1 sample interval +100ppm × reading + 0.6ns) |
>16 averages ± (1 sample interval + 100ppm × reading + 0.4ns) |
|
Sample interval = s/div ÷ 200 | |
Vertical AD Converter | 8-bit resolution, each channel sampled simultaneously |
Vertical VOLTS/DIV Range | 500 μV/div to 10 V/div at input BNC |
Vertical Position Range | 500 μV/div to 20mV/div, ±400 mV |
50 mV/div to 200mV/div, ±2 V | |
500 mV/div to 2V/div, ±40 V | |
5 V/div to 10V/div, ±50 V | |
Vertical Selectable Analog Bandwidth Limit, typical | 20 MHz |
Vertical Low Frequency Response (-3db) | ≤10 Hz at BNC |
Vertical Rise Time at BNC, typical | ≤1.8 ns |
Vertical DC Gain Accuracy | ±3 % for Normal or Average acquisition mode, 10 V/div to 10 mV/div |
±4 % for Normal or Average acquisition mode, 5 mV/div to 500 μV/div | |
Vertical Acquisition Modes | Normal, Peak Detect, Average and HR |
Vertical Acquisition Rate, typical | Up to 2000 waveforms per second per channel (Normal acquisition mode, no measurement) |
Single Sequence | Acquisition Stop Time |
Trigger Mode | Auto, Normal |
Trigger Level | ±4 divisions from center of screen |
0 to 3.3 V | |
Trigger Holdoff Range | 20 ns to 10 s |
Trigger Level Accuracy | 0.2 div × volts/div within ±4 divisions from center of screen |
± (6 % of setting + 40 mV) | |
Edge Trigger Slope | Falling |
Edge Trigger Source | CH1 to CH4 or EXT |