Description
Technical specifications
Resolution | 0.011 μS/cm or 0.1 μS/cm or 1 μS/cm 0.01/0.1 mS/cm |
Accuracy | Electrode: ±1.0% FS, Instrument: ±1.50% FS |
Temperature Range | (-10 ℃ to 110 ℃) 14 ℉ to 230 ℉ |
Temperature Compensation | (0 ℃ to 100 ℃) 32 ℉ to 212 ℉ |
Temperature Accuracy | 41 ℉ to 140 ℉ (5 ℃ to 60 ℃): 31.1 ℉ to 32.9 ℉ (-0.5 ℃ to +0.5 ℃ ) , Others: 30.2 ℉ to 32.18 ℉ (-1.0 ℃ to +0.1 ℃) |
Electrode Constant | 0.1 cm-1 1 cm-1 or 10 cm-1 |
EC Calibration | 1 point |
Reference Temperature | (25 ℃) 77 ℉, (20 ℃) 68 ℉,(18 ℃) 64.4 ℉ |
DO Measurement Range | (0 mg/L to 20.00 mg/L) (ppm) (0% to 200.0%) |
DO Resolution | 0.1 mg/L or 0.01 mg/L (ppm), 1% or 0.1 % |
DO Accuracy | ± 0.40 mg/L, ± 0.3% |
Response Time | ≤ 30 s (25 ℃) 77 ℉, (90% response) |
Residual Current | ≤ 0.1 mg/L |
Salinity Compensation | (0 ppt or 45 ppt)(automatic) |
Pressure Compensation | (80 Kpa to 105 kPa) 11.6 PSI to 15.2 PSI (automatic) |
Automatic Calibration | full oxygen in air and zero air calibration |
Electrode Type | Polarographic |
Data Storage | 300 sets |
Power Supply | DC 12 V/1 A (with power adapter) |
Communications | USB, Bluetooth |
Storage Contents | Series number, measuring value measuring unit and temperature |
Ambient Temperature | (5 ℃ to 35 ℃) 41 ℉ to 95 ℉ |
Ambient Humidity | ≤ 85% |
Dimensions | (240 mm × 170 mm × 70 mm) 9.45 in × 6.69 in × 2.76 inTop of Form |
Weight | 1.32 lb (600 g) |