Technical Specifications
AC Output Voltage Range | 300 V to 5000 V |
AC Voltage Accuracy | ± (1% readings + 5 V) |
AC Breakdown Current | 0 mA to 100.00 mA, short time 200 mA peak value |
AC Voltage Frequency | 50 Hz or 60 Hz |
DC Output Voltage Range | 300 V to 6000 V DC |
DC Voltage Resolution | 1 V |
DC Voltage Accuracy | ± (1% readings + 5 V) |
Insulation Resistance | 1 MΩ to 2000 MΩ |
Ground Resistance | 10 mΩ to 600.0 mΩ (voltage not exceeding 12 V) |
AC Output Voltage | Range: 300 V to 5000 V AC Resolution: 1 V Accuracy: ± (1% readings + 5 V) |
AC Output Frequency | Range: 50 Hz or 60 Hz |
AC Breakdown Current | Range: 0 mA to 100.00 mA, short time 200 mA peak value Resolution: 0.001 mA or 0.01 mA Accuracy: ± (1% + 0.002 mA) |
AC Ramp Up Line | Range: 0 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
AC Slow Down Time | Range: 0 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
AC Testing Time | Range: 0.5 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
AC Current Compensation | Range: 0 mA to 99.999 mA Resolution: 0.001 mA or 0.01 mA |
AC Arc Detection | Range: 0 mA to 100.00 mA Resolution: 0.01 mA Accuracy: ±5% |
DC Output Voltage | Range: 300 V to 6000 V DC Resolution: 1 V Accuracy: ± (1% readings + 5 V) |
DC Breakdown Current | Range: 0 mA to 9.999 mA Resolution: 0.001 mA Accuracy: ± (1% + 0.002 mA) |
DC Ramp Up Time | Range: 0 s to 999.9 s Resolution: 0.01 s Accuracy: ± (1% + 0.1 s) |
DC Slow Down Time | Range: 0 s to 999.9 s Resolution: 0.01 s Accuracy: ± (1% + 0.1 s) |
DC Testing Time | Range: 0.5 s to 999.9 s Resolution: 0.01 s Accuracy: ± (1% + 0.1 s) |
Insulation Resistance (Insulation Resistance Test) | Range: 0.5 s to 999 GΩ Resolution: 0.01 M or 0.1 M or 1 M or 10 M or 100 M or 1 G Accuracy: 0.60 M to 9.99 M ± (1% readings + 0.03 M) 10.0 MΩ to 99.9 MΩ ± (3% readings + 0.3 M) 100 MΩ to 9999 MΩ ± (5% readings + 3 M) 10.0 GΩ to 100 GΩ ± (15% readings + 0.8 G) |
Ramp Up Time (Insulation Resistance Test) | Range: 0 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
Slow Down Time (Insulation Resistance Test) | Range: 0 s to 999.9 s Resolution: 0.01 s Accuracy: ± (1% + 0.1 s) |
Testing Time (Insulation Resistance Test) | Range: 0.5 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
Delay Time (Insulation Resistance Test) | Range: 0 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
Output Current (Ground Resistance Test) | Range: 1.0 A to 32.0 A (optional 42 A) Resolution: 0.1 A Accuracy: ± (1% readings + 0.2 A) |
Output Frequency (Ground Resistance Test) | Range: 50 Hz or 60 Hz |
Ground Resistance (Ground Resistance Test) | Range: 10 mΩ to 600.0 mΩ (voltage not exceeding 12 V) Resolution: 0.1 mΩ Accuracy: ± (1% readings + 2.0 mΩ) |
Resistance Compensation (Ground Resistance Test) | Range: 0 mΩ to 600.0 mΩ Resolution: 0.1 mΩ Accuracy: ± (1% readings + 2.0 mΩ) |
Testing Time (Ground Resistance Test) | Range: 0.5 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
Output Voltage (Leakage Current Test) | Range: 30.0 V to 300.0 V Resolution: 0.1 V Accuracy: ± (0.2% readings + 0.5 V) |
Output Current (Leakage Current Test) | Range: 16.000 A Resolution: 0.001 A Accuracy: ± (0.2% + 0.01 A) |
Leakage Current (Leakage Current Test) | Range: 0.0050 mA to 12.000 mA Resolution: 0.0001 mA or 0.001 mA Accuracy: ± (1% + 3 uA) |
Testing Time (Leakage Current Test) | Range: 1.0 s to 999.9 s Resolution: 0.1 s Accuracy: ± (1% + 0.1 s) |
Human Network (Leakage Current Test) | Range: 6 kinds |