GAOTek Turbidity Meter

GAOTek turbidity meter combines the scattered light method and the transmitted light method and uses a near-infrared light source to eliminate chromaticity interference.

SKU: GAOTek-PLM-136 Category:
Description

Description

Overview

GAOTek turbidity meter combines the scattered light method and the transmitted light method and uses a near-infrared light source to eliminate chromaticity interference. Two detectors, 90°and 180°, are set in the instrument. The 180°detector receives the transmitted light and the 90°detector receives the scattered light. Measure the ratio of the intensity of scattered light and transmitted light and calculate the concentration of suspended solids and turbidity in the sample.

 

Technical Specifications

Range 0 to 20 NTU Turbidity
0 to 1000 NTU Turbidity
0 to 1000 mg or L TSS
Indication error ±0.1 NTU or ±5% NTU (0 to 20NTU Turbidity)
±0.3 NTU or ±8% NTU (0 to 1000NTU Turbidity)
±0.3 mg/L or ±8% mg/L(0 to1000mg/L  Tss)
Repeatability <5NTU, ≤±0.05NTU≥5NTU,≤±1%NTU 0 to 20Turbidity
<10NTU, ≤±0.1NTU≥10NTU, ≤±1%NTU 0-1000Turbid
<10mg or L, ≤±0.1mg/L≥10mg/L, ≤±1%mg or L 0-1000TSS
Calibration 4 points calibration
(optional calibration solution) turbidity
Zero drift ≤±0.1 NTU
Working temperature -10 to  55 ℃
Power lithium battery(Optional dry battery)
Size 170*72*44 mm
Weight 290 (g)