Measurement Frequency |
100 Hz to 100 kHz |
Basic Measurement Accuracy |
0.15 % |
Equivalent Circuit |
Series, parallel |
Mathematical Functions |
Percentage deviation |
Range Way |
Automatic, hold, manual selection |
Trigger Mode |
Automatic, manual, external, bus |
Measure (>= 1 KHz) |
Fast 30 times per second, middle speed 10 times per second, low speed 3 times per second |
Average Time |
1 to 255 |
Delay Time |
0 sec to 6 sec, step by 1 ms |
Calibrating |
Open or short circuit or quick reset |
Display Mode |
Direct reading, delta %, V or I |
Displayer |
5-digit display of main and minor parameters, 4.3 in (10.9 cm) true color LCD displayer |
Output Impedance |
30 Ω and 100 Ω |
Display Of |Z|, R, X, ESR |
0.1 mΩ to 99.999 MΩ |
Display Of C |
0.01 pF to 9.9999 F |
Display Of L |
0.01 µH to 99999 H |
Display Of D |
0.0001 to 9.9999 |
Display Of Q |
0.0001 to 9999.9 |
Degree |
-179.999 degrees to 179.999 degrees |
Radians |
-3.1416 to 3.1416 |
Delta % |
-999.99 % to 999.99 % |
Comparator Function |
5 grade of sorting function |
Storage |
More than 100 sets of instrument settings for storage or call, U disk extension of more than 500 sets |
Interface |
RS232C, HANDLER, USB and HOST standard |
Dimensions |
10.4 in x 3.9 in x 13.3 in (265 mm x 100 mm x 340 mm) |
Weight |
7.7 lb (3.5 kg) |