Description
Overview
GAO Tek multi-gas analyser is used for portable and rapid detection of toxic and harmful gas concentration, temperature and humidity measurement and exceeding the standard alarm. a 2.5-inch high-definition colour screen is used to display the concentration in real time, and gas. The main detection principles are electrochemical, infrared, catalytic combustion, Thermal conductivity, PID photo ion.
Features
- Waterproof, dustproof, explosion-proof, shockproof, and anti-static
- Quick response with >10m sampling distance
- Simultaneously detects 1-5 gases with flexible unit options
- Offers three display modes, including real-time curves
- Enhances linearity and accuracy
- Automatic error identification, reducing human errors
Technical Specifications
Detection range | 0 ppm to 1 ppm, 10 ppm, 100 ppm, 1000 ppm, 5000 ppm, 50000 ppm, 100000 ppm, 200 mg/L, 100% LEL, 20%, 50%, 100%Vol, can be selected; and other ranges can be customised. |
Resolution | 0.01ppm or 0.001ppm (0 ppm to 10 ppm);
0.01ppm (0 ppm to 100 ppm), 0.1ppm (0 ppm to 1000 ppm), 1ppm (0 ppm to 10000 ppm or more), 0.01 mg/l (0 ppm to 200 mg/l), 0.1% LEL, 0.01%, 0.001% Vol |
Detection principle | Electrochemical, catalytic combustion, infrared, thermal conductivity, PID photoionization, and so on, depending on the type of
gas, range, field environment and user demand. |
Sensor service life | Electrochemical principle: 2 years to 3 years; Oxygen gas: 2 years or 6 years can be selected; infrared principle: 5 years to 10 years; catalytic combustion: 3 years; thermal conductivity: 5 years; PID photoionization: 2 years to 3 years. |
Allowable error | ≤±2% F.S (other or higher precision levels can be customised) |
Linearity | ≤±2% |
Repeatability | ≤±2% |
Uncertainty | ≤±2% |
Response time | T 90 ≤ 20 s |
Recovery time | ≤30 s |
Power supply | 3.7 VDC 6000 mAh high-capacity rechargeable polymer battery |
Display mode | 2.5-inch high-definition colour screen |
Detection mode | Built-in pump-suction type measurement, and flow rate at 500 ml / min. Calibration flow rate should be greater than 500 ml / min, three-way pipe should be connected, so as to ensure that the excessive gas will be discharged from the bypass |
Alarm mode | It is possible to set a sound & light alarm, vibration alarm, sound & light + vibration alarm, and turn off the alarm. |
Communication interface | USB (charging and communication), optional: RS232, infrared communication, automatic recognition |
Data storage | Standard configuration is the data storage capacity for 100, 000 entries |
Explosion-proof type | Intrinsically safe type |
Explosion-proof mark | Exia II CT6 |
Standard accessories | Manual, qualification certificate, warranty card, USB charger (including the data cable), high-grade aluminium instrument case, belt clip, humidity dust filter |
humidity | ≤10% RH to 95% RH |
Protection level | IP 66 |
Working Temperature | -40 °F to 158 °F ( -40 ℃ to + 70 ℃) |
Dimensions | 7.68 in x 3.03 in x 1.81 in (195 mm × 77 mm × 46 mm) |
Weight | 0.77 lb (350 g) |