Description
Overview
GAOTek Fiber Multimeter has an all-in-one loss tester integrated with a stabilized laser source, optical power meter, optical loss tester, ORL tester and length testing functions. The device can perform an automatic bidirectional loss test on a single fiber with a pass or fail assessment. With rich functions, flexible configuration and excellent user-friendly design, the device is ideal for optical loss testing and evaluating the optical fiber link condition in FTTx, LAN and CATV networks.
Features
- All-in-one: SLS + OPM + OLT + ORL + Length test
- Optical power monitoring (Auto power trace)
- Bidirectional loss test on a single fiber
- ORL test
- Pass or fail assessment
- Automatic wavelength identification or switch (AutoID)
- Remote reference value setting
- Internal clock and fiber S/N editable
- 1000 test records (CSV) storage & management
- Data transfer to PC via USBÂ (Driver-free)
- Autotest records saved in local unit or remote unit or both units
- USB power charging
- Over 35Â hrs continuous operation
- Multilanguage support
- No warm-up, quick start
- High-resolution color LCD
- Pocket size, lightweight and easy-to-use
Technical Specifications
Stabilized Laser Source | |||
Wavelength (±20 nm) | 1310 nm or 1550 nm | 1310 or 1490 or 1550 nm | 1310 nm or 1550 nm or 1625 nm |
Spectrum Width | ≤5 nm | ||
Emitter Type | FP to LD | ||
Output Power | >-4 dBm | ||
Power Stability | ±0.05 dB per 0.25 hrs; ±0.10 dB per 8 hrs | ||
Output Mode | CW, 270 Hz, 1 KHz, 2 KHz | ||
Optical Power Meter | |||
Calibrated Wavelength | 850 nm, 1300 nm, 1310 nm, 1490 nm, 1550 nm, 1625 nm | ||
Power Range | -70 dBm to 10 dBm (-60 dBm 10 dBm at 850 nm) | -50 dBm to +27 dBm | |
Display Unit | W per m, W per μ, W per n, W per p, W per dBm per dB (REF) | ||
Detector Type | InGaAs | ||
Accuracy | ±5% ±0.01 nW (±0.5 dBm at 850 nm) | ±5% ±1 nW (±0.5 dB at 850 nm) | |
Resolution | 0.01 dB | ||
Auto Wavelength Identification  | Yes (With Shineway Tech specified Stabilized Laser Source) | ||
Mod Identification | 270 Hz, 1 KHz, 2 KHz | ||
Optical Loss Test | |||
Link Loss Range | 50 dB (1550 nm, 124.27 mi (200 km)) | ||
Link Loss Accuracy | ±0.25 dB | ||
Link Loss Test Time | <2Â seconds per wavelength | ||
ORL Test | |||
ORL Range | 0 dB to 60 dBÂ (APC connector) | ||
ORL Accuracy | ±0.75 dB at (0 dB to 50 dB), ±1.5 dB at (50 dB to 60 dB) | ||
ORL Uncertainty | ±0.5 dB at 20 dB | ||
General Specifications | |||
Connector | FC (Interchangeable SC, ST) | ||
Data Storage | 1000Â records | ||
Data Interface | USB | ||
Backlight Display | Yes | ||
Auto Off | Yes (auto-off after 5 minutes idle) | ||
Power Supply | Lithium Battery or AC Adapter | ||
Battery Life | Continuous operation ≥35 hrs | ||
Operating Temperature | -4Â oF to 122Â oF (-20Â oC to 50Â oC) | ||
Storage Temperature | -40Â oF to 158Â oF (-40Â oC to 70Â oC) | ||
Relative Humidity | 0% to 95% (non-condensing) | ||
Weight | 0.77 lb (350 g) | ||
Dimensions | 6.97 in x 3.15 in x 1.73 in (177 mm × 80 mm × 44mm) |