Description
Overview
GAOTek Process Calibrator with Auto Power Off (TC, RTD signals) with high accuracy and high stability, can be widely used in site calibration of industry instruments and equipment in laboratory. This device has powerful functions, it can simulate to output voltage in millivolt and volt, current in milliampere, and many types of electric signals needed during industry control process can be measured and controlled. It can also test or simulate kinds of TC and RTD signals.
Key Features
- 5 1/2 digit input/output display.
- Manual/auto cold junction compensation and setting.
- The temperature value can be directly measured/output.
- Auto power off when the voltage of battery is low.
- Apply advanced built-in battery management system. 7.4 V/7800 mAh lithium battery, which can easily last for over 8 hours during site measurement work.
- Auto triangular wave/step wave signal output.
- It can store 8 sets of measurement and output data, and recall it to display.
- It has backlight.
- The output has manual/auto step and slope output function. It can arbitrarily set the value of 0 %, 25 %, and 100 % in the range of signal output to test the response speed of transmitter.
Technical Specifications
Functions |
Measurement |
Signal Output |
DC Voltage (Upper of display) | 0—30.000 V | |
DC Current (Upper of display) | 0—24.000 mA | |
Loop Power Supply (Upper of display) | 0—24.000 mA | |
DC V (Bottom) | 0—24.000 V | 0—20.000 V |
DC mV (Bottom) | 0—90.000m V | 0—90.000m V |
DC mA (Bottom) | 0—24.000 mA | Simulation or source output 0—24.000 mA |
Resistance (Bottom) | 0—400.00 Ω
400.0—4000.0 Ω |
5.00—400.00 Ω
400.0—4000.0 Ω |
TC (Bottom) | Type S, R, B, K, N, E, J, T,(mV) | |
RTD (Bottom) | Pt100 (385) Pt100 (3926) Pt100 (3916) Pt200 (385)
Pt500 (385) Pt1000 (385) Cu100 (ITS) Cu50 (ITS) |
|
Other function | Step output, slope output, save and recall data, USB communication, etc. |
Acquainting the Process Calibrator
The keys and ports distribution of calibrator is shown in the following picture.
Top ports / Bottom ports
The following table is the wiring instructions of port function.
Remarks |
Name |
Instructions |
Bottom | V/Ω、COM port | The port to test or output voltage, millivolt, resistance and RTD. |
Bottom | mA+/3 W, mA-/4 W port | 1. The input port to test resistance or RTD third & fourth wire.
2. Positive and negative port to test and output current. |
Top | Thermoelectric couple
micro input/output port+TC- |
Test or output thermoelectric couple port. This port can joint with a micro thermoelectric couple plug with polarity (Flat contactor). The distance from one contactor to the other one is 7.9 millimeter (0.312 inch). |
Top | V/LOOP、COM port | The port to test voltage or provide LOOP for testing current. |
Top | mA、COM port | The port to test current. |
The key of calibrator is shown as follows. The following table is the explanation of their functions.
NO. |
Key Name |
Instructions |
|
1 | V | Choose voltage measurement function, and show the value in the upper part of screen.
Use V/LOOP, COM port to input. |
|
2 | mA | Choose current measurement function, and show the value in the upper part of screen.
Use mA, COM port to input. |
|
3 | LOOP | Start the loop power supply to test current, show the value in the upper
Part of screen. Use V/LOOP, COM port. |
|
4 | CONFIG | Enter into configuration menu of calibrator. Set and calibrate the parameter and show the corresponding parameter in the upper part of screen. | |
5 | EXIT | Exit configuration menu, or exit from the function of saving data or invoking data. | |
6 | V
mA |
Choose voltage function (in the bottom of screen), the screen shows “voltage” correspondingly.
Switch measurement or output through “MEAS” and “SOURCE” keys. Choose mA (current) function (in the bottom of screen), the screen shows “mA” correspondingly. Switch measurement or output through “MEAS” and “SOURCE” keys. Press this key repeatedly and choose source output or simulated output function cyclically. The screen shows “mA SOUR” or “mA SIMU” correspondingly. |
|
mV | Choose millivolt function (in the bottom of screen), the screen shows “mV” correspondingly. Switch measurement or output through “MEAS” and “SOURCE” keys. | ||
TC | Choose TC (thermoelectric couple) function (in the bottom of screen), the screen shows “TC” correspondingly. Switch measurement or output through “MEAS” and “SOURCE” keys. Press this key repeatedly and choose the type of thermoelectric couple cyclically. The screen shows the type of thermoelectric couple correspondingly. | ||
Ω | Choose resistance function. Switch measurement or output through “MEAS” and “SOURCE” keys. | ||
Choose RTC (thermal resistance) function (in the bottom of screen), the screen shows “RTC” correspondingly. Switch measurement or output through “MEAS” and “SOURCE” keys. Press this key repeatedly and choose the type of RTC cyclically. The screen shows the type of thermal resistance correspondingly. | |||
RTD | |||
7 | MEAS | Choose measurement mode, the screen shows “measurement” correspondingly. | |
SOURCE | Choose output mode, the screen shows “output” correspondingly. | ||
To enter data store function, it means to save the measurement value or output value in the bottom of screen into internal address, total 8 addresses.
Press this key repeatedly to store data address. When there is date in the address, this address will be shown bold. |
|||
STORE | |||
RECALL | To enter data recall function, it means to display and output the value saved in the internal address. Press this key repeatedly to select data recall address. | ||
8 | SLOPE STOP | Repeat 0 %-100 %-0 % slope (as 25 % step)
Repeat 0 %-100 %-0 % slope slowly Repeat 0 %-100 %-0 % slope quickly Stop automatic slope of the output function. Stopped for manual output and its manual step function. |
|
9 | Numeric key | Input the digital value in output mode. | |
10 | OUTPUT | In the output mode, input digital value by numeric key and press this key to output the signal from ports. | |
ENTER | When entering into data store function or data recall function, confirm the address to store data or recall data.
And calibrate the time in configuration menu or be used with other keys. |
||
CLEAR | Remove the input data | ||
11 | 0% | Press this key in output mode, it will callout the setting value 0% from the memory and shows output. Press “0%” key and “ENTER” key together, it will save current output volume as 0% value. | |
-25% | Press this key in output mode, it will increase output 25% of the range from 0 % to 100 %. | ||
+25% | Press this key in output mode, it will reduce output 25% of the range from 0 % to 100 %. | ||
100% | Press this key in output mode, it will callout the setting value 100 % from the memory and shows output. Press “100 %” key and “ENTER” key together, it will save current output volume as 100 % value. | ||
12 | Left/right/up/
down direction key |
Press the left or right direction key in resistance measurement function or RTD measurement function, circularly select 2W, 3W, 4W measurement.
Used for calibrating output function and adjusting corresponding functions in the configuration menu. |
|
13 | Wheel | Turning this wheel in voltage, current, mV output function. The value will reduce a unit when rotate it counter-clockwise. The value will increase a unit when rotated clockwise. It does not work in other function gears. | |
14 | CALDN | Used for calibrating the function in the bottom. | |
15 | BL | Back light switch |
Basic Operation
1. Configuration menu function
There are ten items of setting in configuration menu. Press “CONFIG” to enter into. Press EXIT to exit.
Setting 1. Adjusting the time to power-off. (The calibrator will power off after the setting time when there is no any operation.) The factory default value is 20 minutes. The detail operation step and explanation are as follow.
- Press “CONFIG” it shows “power off time XX minutes” in the upper part of screen. The unit is minute.
- Press up or down direction key, to increase or reduce the power-off time, range: 0-60 minutes. When the power-off time is setting as 0, the calibrator will cancel auto power-off. Then it can only be powered off manually.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 2. Adjusting screen contrast. Factory default is 09. The detail operation step and explanation are as follow.
- Press CONFIG, keep pressing and select setting 2. It shows “Screen contrast XX” in the upper part of screen.
- Press up or down direction key, to increase or reduce the number in the screen, range: 0-16. When the number gets larger, the content in the screen gets weaker. When the number gets smaller, the content in the screen gets darker.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 3. Adjusting the delay time of back light (The calibrator will turn off the back light after the setting time when there is no any operation). Factory default value is 5 minutes. The detail operation step and explanation are as follow.
- Press CONFIG, keep pressing and select setting 3. It shows “Turn-off time of back light XX minutes” in the upper part of screen. The unit is minute.
- Press up or down direction key, to increase or reduce the turn-off time of back light, range: 0-20 minutes. When the turn-off time is setting as 0, the calibrator will cancel auto turn-off. Then it can only be turned off manually by pressing BL key.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Attention: It can save the battery and increase the using time to set the turn-off time of back light shorter.
Setting 4. Setting calendar and time. The detail operation step and explanation are as follows.
- Press CONFIG, keep pressing and select setting 4. It shows “Calendar time setting: XX year” in the upper part of screen. The unit is year.
- Press up or down direction key, to increase or reduce year, range: 00-99 year.
- Press ENTER, enter into the setting of month, and continuously set date, hour, minute and second.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 5.Setting temperature unit. The factory default is ℃. . The detail operation steps and explanation are as follows.
- Press CONFIG, keep pressing and select setting 5. It shows “Setting temperature unit: ℃” in the upper part of screen.
- Press up or down direction key, to select ℃ or ℉ circularly. When the temperature unit is changed, the unit in measurement and output TC or RTD function will also be changed.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 6. Setting HART. The factory default is OFF. The detail operation step and explanation are as follows.
- Press CONFIG, keep pressing and select setting 6. It shows “Setting HART: OFF (ON)” in the upper part of screen.
- Press up or down direction key, to select ON or OFF circularly. If select “ON”, select output current (mA) in the bottom of screen and series 250Ωresistance, it shows “HART” in the bottom of screen.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 7.Setting cold junction compensation. The factory default is OFF. The detail operation step and explanation are as follows.
- Press CONFIG, keep pressing and select setting 7. It shows “Cold junction compensation: OFF (ON)” in the upper part of screen.
- Press up or down direction key, to select ON or OFF circularly. If select “ON”, it automatically does cold junction compensation in measurement or output TC function in the bottom of screen. It shows “CJC XX.X”in the bottom of screen, XX.X means the temperature of current cold junction compensation.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 8.Calibrating the temperature of cold junction compensation.
Put the calibrator in the calibration environment for 10 minutes before starting the calibration. Start to calibrate when the temperature inside the instrument is as same as it in the environment. The temperature has better been around ten ℃ .
The environmental measurement thermometer with accuracy ±0.2℃ should be used as standard. The detail operation step and explanation are as follows.
- Press CONFIG, keep pressing and select setting 8. It shows “Cold junction compensation: XX.X℃” in the upper part of screen.
- Press up or down direction key, to increase or reduce the number in the screen. Adjust the temperature in the screen as same as the value in standard environmental measurement thermometer.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 9. System calibration turns on or off. The factory default is (OFF. The detail operation step and explanation are as follows.
- Press CONFIG, keep pressing and select setting 9. It shows “System calibration setting: OFF (ON)” in the upper part of screen.
- Press up or down direction key, to select ON or OFF circularly. If select “ON”, turn on the system calibration. Otherwise, turn off the system calibration.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
Setting 10. Language selection. The detail operation step and explanation are as follow.
- Press CONFIG, keep pressing and select setting 10.
- Press up or down direction key, to select Chinese or English circularly.
- Press CONFIG, enter into the next level setting. Press EXIT to exit configuration menu.
2. Data store and recall.
Data store is to save the measurement or output (analog) data in the bottom of screen. The calibrator can save 8 packs of data in the NVM (nonvolatile memory) for recall in the future. The stored data will not lose after powering off the calibrator. The detail operation step and explanation are as follow.
- Press STORE to enter into data store function. It shows “STORE” and address number “1…8” in the right up corner of screen, total 8 addresses. The current store address number will blink.
- Keep pressing STORE and select store address number circularly. If there is already been data in this address number, it will be shown with black in the screen. If saving data here, the current data will be replaced.
- Press ENTER key, confirm to save data in current selected place. The next data store address number will start to blink.
- Press EXIT, exit the data store function.
Data recall is to display the data in the NVM (non-volatile memory), output with output (analog) data. The detail operation step and explanation are as follow.
- Press RECALL to enter into data recall function. It shows “RECAL” and address number “1…8”of data recall in the right up corner of screen, total 8 addresses.
The current data recall address number will blink.
- Keep pressing RECALL and select data recall address number circularly. If there is already been data in this address number, it will be shown with black in the screen.
- Press ENTER key, confirm to recall the data in current selected place. The next data store address number will start to blink.
4) Press EXIT, exit the data recall function.
3. Back light of screen
Press BL, turn on the back light. It will automatically turn off after the appointed delay time or press BL again to turn it off. The delay time can be set, the default value is 5 minutes.
Measurement functions
A. Measurement functions in the upper part of screen.
1) To measure the voltage of transmitter, please operate as following steps.
① Press V (1), select voltage measurement. The red and black probe connects to V/LOOP and COM port separately.
②Observe the reading in the upper part of screen, like over range “-OL–”.
The maximum measurement value is 30 V. Attention: Do not connect the high voltage over the range to any end of the ports. It may damage the instrument.
2) To measure the current of transmitter, please operate as following steps.
① Press mA (2), select current measurement. The red and black probe connects to mA and COM port separately.
② Observe the reading in the upper part of screen, like over range “-OL–”. The maximum measurement value is 24 mA. Attention: Do not connect the current over the range to any end of the ports. It may damage the instrument.
3) To measure the current with loop power supply LOOP
The LOOP function start a 24 V power supply connects in series with current measurement circuit. It can achieve the function to power up the transmitter and measure the current output of transmitter after the wiring is removed in site. If to use LOOP to measure current, please operate as following steps.
① Press LOOP (3) when the calibrator is in other measurement mode, it shows “LOOP” in the upper part of screen. Then the calibrator provides 24 V loop power supply internally.
② The red and black probe connects to V/LOOP and COM port separately.
Observe the reading in the upper part of screen, like over range “-OL–”.