GAOTek Test Gas Analyzer

GAOTek Test Gas Analyzer adopts imported sensors, which integrates SO2 + SOF2, H2S, CO, H2, HF and moisture content measurement into one gas.

GAOTek-GATI-125

Description

Overview

GAOTek Test Gas Analyzer adopts imported sensors, which integrates SO2 + SOF2, H2S, CO, H2, HF and moisture content measurement into one. gas. This instrument has the characteristics of high precision and stable performance. The instrument uses color LCD English display, imitating Windows operation interface, real-time display of various parameters, fully intelligent operation, massive information storage, built-in rechargeable battery, AC and DC dual use.

 

Features

• Quick 5 min measurement upon startup

• German self-locking connector ensures safe, leak-free operation

• Stores 500 sets of data, 5-inch touch display

• Built-in lithium battery lasts 30 hours continuously

• In-built electronic flow meter for precise measurement

 

 Technical Specifications

Detection amount SO2+SOF2, H2S, HF, H2 ≤ 0.1 µl/L
CO ≤ 0.5 µl/L
Dew point ≤ 32.18 °F (≤ 0.1 ℃) at -4 °F to -76 °F. (-20 ℃ to -60 ℃)
Accuracy Small range SO2 + SOF2, H2S, H2 When the measured value is ≤ 10 µl/ L, the error is ≤ ± 0.3 µl/L
When the measured value is greater than 10 µl/L, the error is ≤ ± 3%
Large range SO2 + SOF2 Error ≤ ± 3%
HF, H2 Error ≤ ± 0.3 µl/L
CO When the measured value is ≤50 µl / L, the error is ≤ ± 2 µl / L
When the measured value is >50 µl / L, the error is ≤ ± 4%
Dew point ≤ ± 0.9 °F (≤ ± 0.5 ℃) at –4 °F to –76 °F (-20 ℃ to -60 ℃)
Linearity SO2+SOF2, H2S, HF, ≥ 99.9%
H2, CO
Dew point ≥ 99.9% when the dew point is –76 °F to 32 °F (-60 ℃ to 0 ℃)
Repetition SO2+SOF2, H2S, HF, H2 When the measured value is ≤10 µl /L, the error is ≤±0.2 µl/L
When the measured value is greater than 10 µl/L, the error is ≤±2%
CO When the measured value is ≤50 µl /L, the error is ≤±1 µl/L
When the measured value is >50 µl /L, the error is ≤±2%
Dew point ≤ ± 0.9 °F (≤±0.5 ℃)
Quantity SO2+SOF2 0 µl/L to 100 µl/L, optional: 0 µl / L to 2000 µl / L
H2S, H2 0 µl / L to 100 µl/L
HF 0 µl / L to 20 µl/L
CO 0 µl / L to 500 µl/L
Dew point -148 °F to +68 °F. (-100 ℃ to +20 ℃)
Gas flow Decomposition products 200 ± 20 ml/min, flow accuracy ≤±3%
Dew point 500 ±50 ml/Min
Environmental conditions Temperature -4 °F to +122 °F (-20 ℃ to +50 ℃)
Humidity 10% RH to 95% RH at 68 °F (20 ℃), no condensation
Altitude ≤ 4000 m
Communication USB
Intake pressure range 21.76 PSI to 145.04 PSI (0.15 MPa to 1.0 MPa)
Power supply AC / DC dual-use: AC 110 to 220 ± 10% v, 50 to 60 ± 1 hz; built-in 18 MH large-capacity high-performance lithium battery, which can work continuously for more than 30 h when fully charged, with prompts for displaying power and voltage too high, undervoltage and protective function