GAOTek push electrical switch is an online water quality monitoring instrument with a microprocessor. This instrument is equipped with conductivity electrodes of different types and constants for continuous monitoring and control of the conductivity and temperature values of aqueous solutions. Widely used in fields such as power plants, petrochemicals, metallurgy, paper industry, environmental water treatment, light industry electronics, etc. Monitoring and control of raw and produced water quality for water production equipment such as power plant cooling water, make up water, saturated water, condensate and boiler water, ion exchange, reverse osmosis EDL, and seawater distillation. The instrument adopts an LCD display screen; Intelligent Chinese menu operation; Equipped with 2 current outputs, freely set measurement range, high and low limit alarm prompts, two sets of relay control switches, adjustable hysteresis range, and automatic cleaning switch settings; Automatic or manual temperature compensation; Multiple automatic calibration methods for electrodes. According to user needs, electrodes can be installed in various forms such as flow through, immersion, pipeline, and flange.
Features
- Design based customization
- Sample based customization
- Raw material traceability identification
- Finished product inspection
Technical Specifications
Measurement Range | 0.01 to 20 μ S per cm (with 0.01 electrode) 20 to 200 μ S per cm (with 0.1 electrode) 200 to 2000 μ S per cm (with 1.0 electrode) 2000 to 20000 μ S per cm (equipped with 10.0 electrodes) 6000 to 60000 μ S per cm (with 30.0 electrodes) |
Power Supply | 220 V AC ± 10 %, 50 ± 1 Hz, power ≤ 3 W 24 V DC, power ≤ 1 W (to be predetermined), 12 V DC, power ≤ 1 W (to be predetermined) |
Stability | ± 0.2 % FS ± 1 word per 24 H |
Basic Error Of Electronic Unit | Conductivity: ± 0.5 % FS |
Signal Output | 4 to 20 mA (load resistance < 750 Ω); RS485 (optional) |
Three Sets Of Relay Control Contacts | 3 A 240 V AC, 6 A 28 V DC or 120 V AC |
Instrument Basic Error | Conductivity ± 3.0 % FS, Temperature ± 32.90 °F (± 0.5 ℃), 0 °F to 140.00 °F (0 ℃ to 60 ℃) |
Work Environment | Environmental temperature: – 50 °F to 140 °F (-10 to 60 ℃) Relative humidity: not more than 90 % |
Automatic Temperature Compensation Range | 0 °F to 211.82 °F (0 ℃ to 99.9 ℃), with 77 °F (25 ℃) as the reference temperature |
Overall Size | 3.78 in x 3.78 in x 5.51 in (96 mm× 96 mm× 140 mm) |